Adres: 1264. SOKAK (ESKİ 367) 4
Tel 1: 0 (312) 395 13-27
Faks:
Email 1: info@med-san.com
Web: https://www.med-san.com
BNT-2000 Radiant Heater
● Temperature Control Mode: preheat control, servo control and manual control;
● Instant Parameter Display: set temperature, skin temperature, heat power ratio;
● Built-in APGAR Timer and >37℃ preset function
● Audible / Visual Alarms: temperature deviation, sensor failure, power failure, temperature override;
● Rotating Radiant Head Tilt Bed;
● X-ray cassette tray under baby bed, high illumination examination light;
● Use anti-explosion quartz infrared tube as radiant heating source;
TECHNICIAL SPECIFICATIONS:
Temperature Control Range: 25 37 -37 ℃,
Temperature override mode: 38℃
Temperature sensor accuracy: ≤0.3℃
Bed surface temperature uniformity: .02.0 ℃
Temperature rise time: ≤45mins
Heat output indicates: 0-100% at 10%
Radiant box vanishing angle adjustment range: 0-90° steplessly adjustable
Bed tilt adjustment: adjustable at ±10°
Bed Tray Size: 1700px * 1350px
Bed tray Max. load: 10kg
Transfusion pole Max. load: 2kg
Rack Max. load: 2kg
ALARM
Sensor failure alarm
Over temperature alarm
Deviation alarm: ±1℃
power failure alarm
APGAR time: audible alarm when running from 50 seconds to 1 minute, 4 minutes 50 seconds to 5 minutes,
9 minutes 50 seconds to 10 minutes,
ENVIRONMENTAL TEMPERATURE
(Please do not use it outside the permitted environment.)
Operating temperature range: 18 ℃ ~ 30 ℃
Store temperature range: -40℃~+55℃
ENVIRONMENTAL HUMIDITY
Operating humidity range: 30%RH ~ 75%RH
Store humidity range: ≤93% RH
ATMOSPHERIC PRESSURE
Transport and Storage atmospheric pressure range: 500hPa ~ 1060hPa
Operate the atmospheric pressure range: 700hPa ~ 1060hPa
FLOW RATE
Media medium flow rate: ≤0.3m/s